An adsorber (active charcoal) removes remaining oil vapors before the pure helium is sent back to the cold head. Maintenance and Servicing
Operates best between 5°C and 35°C (41–95°F). Capacity may degrade by 5% if temperatures exceed 28°C. sumitomo f50 compressor manual new
Low-pressure helium returns from the cold head through a return connector. An adsorber (active charcoal) removes remaining oil vapors
The gas is pressurized within the compressor capsule, generating heat. Low-pressure helium returns from the cold head through
Requires a flow rate of 7–10 L/min (1.8–2.6 gal./min) at an inlet temperature of 4–28°C. Do not use demineralized water for cooling.
The is a critical component in cryogenic systems, primarily used to power Gifford-McMahon (GM) cryocoolers and pulse tubes . Part of SHI's Freedom® series, this mid-size, water-cooled unit is designed for high-reliability applications such as MRI magnets and laboratory cryostats. Technical Specifications
The unit should never be tilted more than 30° during transport or operation.